郑小玲, 邹芸, 杨昌林, 唐柏彬, 张舸, 李建平. 辣椒油中罗丹明B的测定能力验证[J]. 食品工业科技, 2017, (23): 188-191. DOI: 10.13386/j.issn1002-0306.2017.23.035
引用本文: 郑小玲, 邹芸, 杨昌林, 唐柏彬, 张舸, 李建平. 辣椒油中罗丹明B的测定能力验证[J]. 食品工业科技, 2017, (23): 188-191. DOI: 10.13386/j.issn1002-0306.2017.23.035
ZHENG Xiao-ling, ZOU Yun, YANG Chang-lin, TANG Bo-bin, ZHANG Ge, LI Jian-ping. Validation of proficiency test of Rhodamine B in chili oil[J]. Science and Technology of Food Industry, 2017, (23): 188-191. DOI: 10.13386/j.issn1002-0306.2017.23.035
Citation: ZHENG Xiao-ling, ZOU Yun, YANG Chang-lin, TANG Bo-bin, ZHANG Ge, LI Jian-ping. Validation of proficiency test of Rhodamine B in chili oil[J]. Science and Technology of Food Industry, 2017, (23): 188-191. DOI: 10.13386/j.issn1002-0306.2017.23.035

辣椒油中罗丹明B的测定能力验证

Validation of proficiency test of Rhodamine B in chili oil

  • 摘要: 为了解我国食品检测实验室对罗丹明B检测的整体水平,同时验证和加强相关实验室的检测能力,国家认证认可监督管理委员会(CNCA)组织,重庆出入境检验检疫局技术中心具体协调与实施辣椒油中罗丹明B测定能力验证工作。遵循GB/T 27043-2012(等同ISO/IEC导则17043:2010)、GB/T 28043-2011、CNAS-GL02:2014、CNASGL03:2006等相关规定,实施和评价此次能力验证。采用F检验方法进行样品的均匀性检验,采用t检验方法进行稳定性检验,采用Z比分数评价参与实验室的测试结果。134家实验室参加能力验证测试,其中124家实验室结果为满意,满意率为92.5%,8家实验结果为不满意,1家实验室由于设备故障中途退出本次能力验证。参加能力验证的绝大多数实验室可以准确检测罗丹明B,我国相关检测实验室具有较高水平。 

     

    Abstract: In order to find out the capability of Rhodamine B detection in China's food inspection laboratories, meanwhile verify and strengthen the capability of Rhodamine B determination, the National Certification and Accreditation Supervision and Management Committee ( CNCA) organized proficiency test ( PT) of detection of Rhodamine B in Chili oil.Chongqing Entry-Exit Inspection & Quarantine Bureau Technology Center is responsible for specific coordination and implementation.This assessment was performed according to GB/T 27043-2012 ( equivalent to ISO/IEC guideline 17043: 2010) , GB/T 28043-2011, CNASGL02: 2014, CNAS-GL03: 2006, and the international harmonized protocol for proficiency test.The homogeneity of samples was assessed by F test, and the stability was assessed by t-test.The median was used as assigned value to calculate a z-score which was used as a numerical indicator to show participants ' performance in the program. 134 laboratories participated this proficiency test, 124 laboratories were observed as satisfactory, the satisfaction rate was 92.5%, and 8 laboratories showed unsatisfactory result, 1 laboratory quitted due to the equipment's failure.In conclusion, most laboratories taking part in the PT have good competence in analyzing Rhodamine B and most of domestic related laboratories possess a high level of detection.

     

/